The pump-rate calibration was done by pumping against a fixed pressure of 1200 dbars. This was accomplished using a pressure relief valve to vent water as necessary in order to keep the pressure from rising above 1200 dbars. By measuring the time required to pump a fixed volume (ie., 4 A/D counts) then the pump-rate at a given voltage is known. A lab power supply was used to supply current at a voltage that was fixed while each pump-rate measurement was made.
Least-squares methods were used to fit the model of the pump motor (see
equation (3)) to the data from Table 3.
The optimal parameter values
and
yielded an RMS
misfit of only 0.31 V. The right-most two columns of
Table 3 show the predicted voltage,
,
and
voltage error,
.
The pump-rate is a known and
well-behaved function of pressure (ie., current) and supply voltage.