Float 1759 has experienced the Druck pressure sensor (SnowFlake) problem. This float has SBE41 SerNo 1531 with Druck pressure sensor SerNo 5853. The pressure sensor was last calibrated in April, 2004. Both the serial number and the date are well beyond what are supposed to be thresholds for obtaining pressure sensors meeting the 'gold standard'.